IEC 60601-1 is the international safety standard for medical electrical equipment. It sets requirements for basic safety and essential performance, covering electric shock, leakage current, insulation, temperature, fire and mechanical hazards. EN 60601-1 is the European adoption of the same standard, and related parts such as IEC 60601-1-2 cover electromagnetic compatibility.

01

What IEC 60601-1 covers

IEC 60601-1 is the international safety standard for medical electrical equipment, published by the International Electrotechnical Commission. Its full title refers to general requirements for basic safety and essential performance. Basic safety means protection from hazards such as electric shock, excessive leakage current, overheating, fire and moving parts. Essential performance means the clinical functions that must keep working, or fail safely, for the patient to be safe.

The standard applies to equipment that diagnoses, treats or monitors a patient and has an electrical connection, from infusion pumps and patient monitors to surgical displays and cart-mounted computers. It also covers medical electrical systems, where several devices, some of them ordinary IT equipment, are combined into one installation.

EN 60601-1 is the European adoption of the same standard, harmonised under the EU Medical Device Regulation. In the US it is published as ANSI/AAMI ES60601-1, and in Canada as CSA C22.2 No. 60601-1. The current edition is the third, published in 2005 and amended in 2012 and 2020.

02

Main requirements of the standard

IEC 60601-1 is a long document, but most of its requirements fall into a handful of areas.

  • //
    Leakage current. Strict limits on current that could flow through a patient or operator.
  • //
    Insulation and isolation. Defined levels of protection, measured as MOOP and MOPP.
  • //
    Temperature and fire. Limits on surface temperatures, especially for parts that touch the patient, and on materials and enclosures to contain a fire.
  • //
    Mechanical safety. Stability, moving parts, sharp edges and the strength of mounts, carts and handles.
  • //
    Risk management. Since the third edition, manufacturers must run a risk management process to ISO 14971, and many requirements refer back to it.
  • //
    Power supplies. Medical grade supplies help a system meet the limits.
03

Leakage current limits

Leakage current is the small current that flows from mains-powered equipment through insulation, earth or a person. The limits depend on the path and on the type of applied part, the part that touches the patient. NC is normal condition and SFC is single fault condition.

Current Type B applied part Type BF applied part Type CF applied part
Earth leakage, NC / SFC 5 mA / 10 mA 5 mA / 10 mA 5 mA / 10 mA
Touch current, NC / SFC 100 µA / 500 µA 100 µA / 500 µA 100 µA / 500 µA
Patient leakage AC, NC / SFC 100 µA / 500 µA 100 µA / 500 µA 10 µA / 50 µA

Type B parts are not isolated from earth, Type BF parts are floating, and Type CF parts are floating and suitable for direct contact with the heart. A device’s label shows its applied part type with a symbol of a figure, a figure in a box or a heart.

FAQ · 01
What is the difference between MOOP and MOPP?
Both are means of protection under IEC 60601-1. MOOP protects operators and MOPP protects patients, with higher test voltages and larger creepage distances. Equipment in contact with or near a patient typically needs 2 MOPP.
Read the full FAQ —
04

The patient environment and medical systems

The standard matters most near the patient. Equipment within the patient environment, the space around a patient where contact is possible, has to meet IEC 60601-1 itself or be combined with other equipment into a compliant medical electrical system, for example by using isolation or a medical grade power supply. The patient environment is usually taken as 1.5 m around the patient and up to 2.5 m above the floor.

Clause 16 of the standard covers medical electrical systems. A cart or workstation is assessed as a whole, not just its parts, so a computer, display, capture card and isolation transformer must together meet the leakage and protection limits. Our guide to the medical grade power supply explains how isolation is usually achieved.

05

The IEC 60601 family

IEC 60601-1 is the general standard. It sits at the top of a family of related documents that add requirements for specific risks or specific devices.

  • //
    Collateral standards. Numbered IEC 60601-1-x, they add requirements that apply across many devices: -1-2 for electromagnetic compatibility, -1-6 for usability, -1-8 for alarm systems and -1-11 for home healthcare use.
  • //
    Particular standards. Numbered IEC 60601-2-x, they cover specific device types such as surgical equipment, patient monitors or ultrasound, and can override parts of the general standard.
  • //
    Electromagnetic compatibility. IEC 60601-1-2 sets emission and immunity limits, so a device neither disturbs nor is disturbed by other equipment, including radios and phones nearby.
06

UL 60601-1 and ANSI/AAMI ES60601-1 in the US

In the United States, IEC 60601-1 is adopted as ANSI/AAMI ES60601-1, published by the Association for the Advancement of Medical Instrumentation. It carries the IEC text with US national differences, mainly for mains wiring, plugs and markings that suit the US electrical system, and the FDA recognises it as a consensus standard for device submissions. Canada uses the equivalent CSA C22.2 No. 60601-1.

UL 60601-1 was the earlier US version, based on the second edition of IEC 60601-1. It was withdrawn when the US moved to the third edition, with the FDA transition ending around 2013, so new equipment is assessed to ES60601-1 instead. The name lives on because test houses such as UL Solutions, Intertek and CSA Group certify equipment to ES60601-1, and datasheets often shorten that to “UL 60601-1”.

Region Standard Based on Notes
International IEC 60601-1 Edition 3.2 (2005, amended 2012 and 2020) CB Scheme test reports accepted by many national bodies
European Union EN 60601-1 IEC 60601-1 Harmonised under the EU Medical Device Regulation
United Kingdom BS EN 60601-1 EN 60601-1 Used for UKCA and CE conformity
United States ANSI/AAMI ES60601-1 IEC 60601-1 with US differences FDA-recognised; certified by test houses such as UL Solutions and Intertek
Canada CSA C22.2 No. 60601-1 IEC 60601-1 with Canadian differences Used for Health Canada licensing
US (legacy) UL 60601-1 IEC 60601-1 second edition Withdrawn; replaced by ES60601-1

For equipment sold on both sides of the Atlantic, the usual route is one CB test report to IEC 60601-1 that also covers the US and Canadian national differences, rather than separate test programmes for each market.

07

What it means for medical computers

Most computers and monitors used in hospitals are not medical devices themselves, but they often sit in the patient environment or connect to devices that are. A medical-grade computer or display built to IEC 60601-1 removes the need for extra isolation and makes the system assessment simpler. A standard office PC can still be used in some places, but usually only outside the patient environment or behind an isolation transformer.

When specifying, ask for the certificate and the CB test report, check which edition and amendments it covers, and confirm the IEC 60601-1-2 EMC test level. The medical cart computer case study used EN 60601-1 certified hardware, and our explainer on the medical cart covers how those systems are built.

08

Quick answers on IEC 60601-1

IEC 60601-1: common questions

What is IEC 60601?

A family of medical electrical safety standards. IEC 60601-1 is the general standard, with collateral standards numbered 60601-1-x and particular standards for specific devices numbered 60601-2-x.

What is the difference between IEC 60601-1 and EN 60601-1?

Publisher, not content. EN 60601-1 is the European adoption of the IEC standard, harmonised under EU medical device law, with essentially the same technical requirements.

What edition of IEC 60601-1 is current?

The third edition with amendments. Edition 3 was published in 2005, amended in 2012 (edition 3.1) and again in 2020 (edition 3.2).

Does IEC 60601-1 cover EMC?

Through IEC 60601-1-2. Electromagnetic compatibility sits in the collateral standard IEC 60601-1-2, which is normally tested alongside the general standard.

Does a hospital computer need IEC 60601-1?

If it is in the patient environment, usually. Inside the patient environment it must meet the standard itself or be made compliant as part of a medical electrical system.

What is UL 60601-1?

The former US version of the standard. UL 60601-1 was based on the second edition of IEC 60601-1 and has been replaced by ANSI/AAMI ES60601-1, which follows the third edition. Equipment described as “UL 60601-1” today is normally certified to ES60601-1.

Our medical grade monitors and medical panel PCs are built to IEC 60601-1, and we have supplied them across medical and pharmaceutical projects. We integrate and validate each medical system as standard, with the isolation and power chosen for where it will sit.